Double Pumps LCD Touch Screen Hemodialysis Machine
Apply to hemodialysis, hemoperfusion, ISO UF, plasma exchange, etc.
User friendly operation
All-around security system
Precise UF system
Powerful functions
User friendly operation
10” LCD touch screen
Automatically activates treatment program when blood is detected flowing through
Auto self-test, prime, rinse, disinfection, hot disinfection, treatment and power off, etc.
Records in table and graphs
All-around security system
Conductivity
Temperature
Air bubble & Blood level
Blood leakage
Blood pressure
Arterial, Venous & TM Pressure
Anti-coagulation
Backup battery
Precise UF system
Classic waterway system controls the UF precisely
Double circulation waterway system to prevent cross-infection
Powerful functions
Na+, UF & HCO3- Profiles
Suitable for any dialysate formula
Dialysate flow linearly adjustable
Kt/V
Optional: Bicarb-cartridge holder, Pyrogen filter port, Network connection, etc.
Other features
• Touch screen was equipped and it can show 7 Na and UF profiles
• Backup power supply, Kt/V, HCO3¯ adjustments, single needle, acetate treatments
• Input water pressure from 0-0.6Mpa
• Dry powder dialysis function
1.Volume (length × width ×height): 370mm×340mm ×1570mm
2. Weight: about 90kg
3. Power supply voltage: AC220V/110V 50 HZ / 60 HZ ,Power: 1500W
4. Blood pump/spare pump: Flux: 15 ~ 340ml/min (D 6mm) 20 ~ 460 ml/min (D 8mm)
5. Heparin pump: Flux: 0ml/h~10ml/h (±5%) Precise: 0.1ml/h
10ml/20ml/30ml (selectable)
6. Arterial pressure: display scope:-300mmHg~+300mmHg (±10mmHg)
7. Venous pressure: display scope:-50mmHg~+300mmHg (±10mmHg)
8. Dialysate flow:
300ml/min~800ml/min linearity adjustable (+10%) (-5%)
Temperature: 35.0ºC~39.0ºC
Resolving rate: 0.1ºC
Conductivity:13mS/cm~15.5mS/cm (±0.1 mS/cm)
9. UF flow scope: 0~1800ml/h (±30ml/h ) 0~4000ml/h(optional)
10. ISO UF flow scope: 0~2000ml/h (±30ml/h)
11. TMP: Scope:-100mmHg~+600mmHg (±20mmHg)
12. Blood leakage monitor: over 1ml blood per liter dialysate (flow:500ml/min)
13. Blood level monitor: ultrasonic sensor
14. Air bubble monitor: Infrared and response threshold value: Single air bubble of 200 ml exists when Blood flux is 200ml/min
15. Inflow pressure: 0 MPa~0.6MPa
16. Inflow temperature: 5ºC~30ºC
17. Environment temperature: 10ºC~30ºC,Relative humidity≤70%
18. Rinse/disinfection: chemical disinfection citric acid , Peracetic acid and oxalic acid)
19. Hot rinse: 85ºC
20. Back-up power supply: last for 15-30min after electric-cut.
Clinical Application
Hemodialysis Therapy
Pure ultrafiltration Therapy
Hemoperfusion Therapy
Plasma exchange Therapy
Na and UF profiling:
Be applicable for reducing the cases during dialysis treatment such as low/high pressure, muscle convulsion, headache, nausea, vomit etc.
Kt/V function:
The sufficiency of dialysis can be reflected by Kt/V.
HCO3 adjustment function